Vertilon2018-09-14T12:06:13+00:00

Vertilon
PhotoniQ

KETEK’s PA3325-WB-0808 SiPM Array is fully compatible with Vertilon’s PhotoniQ, a series of high performance multichannel data acquisition systems that interface to optical sensors, collect and process their charge signals, and output the resulting data over a USB connection to a PC. With the included graphical user interface the data can be displayed and logged or further processed by custom applications.
Vertilon’s PhotoniQ data acquisition systems are available in various speed and channel configurations. Together with KETEK’s SiPM array it can be used e.g. in radiation detection systems, medical imaging, biophotonics, high energy physics – or just for the evaluation of our SiPM arrays.

PM3325-WB Energy Spectrum with Vertilon DAQ

Energy Spectrum

 
Figure 1: Energy spectrum obtained with a ²² Na source and an LYSO crystal of 3 x3 x 20 mm³ on a PM3325-WB SiPM at 4.5 V. The charge signal is integrated in a time window of 350 ns. The obtained energy resolution at 511 keV is 12.6 % after correction for the SiPM non-linearity.

Energy Spectrum

Figure 2: Energy spectrum obtained with a ²² Na source and an LYSO crystal of 3 x3 x 5 mm³ on a PM3315-WB SiPM at 4.5 V. The charge signal is integrated in a time window of 350 ns. The obtained energy resolution at 511 keV is 13.9 %

PM3315-WB Energy Spectrum with Vertilon DAQ