KEY PARAMETERS VITUS H30 | ||
| Class | Premium | Standard |
| Energy Resolution | < 133 eV | < 144 eV |
| Peak to Background | > 10000 | > 6000 |
| Peak to Tail | > 2000 | > 1000 |
| Absorption Depth | 450 µm Si | |
| Peak Shift Stability up to 100 kcps | < 1eV | |
| Max. input countrate | 500.000 cps | |
| Windows | 8µm or 12.5 µm Be | |
| Cooling Performance | ∆T > 70 K | |
| Collimator Materials | Zr - Pd - Mo | |
REGISTER HERE AND GET THE DETAILED | ||
The energy resolution of the VITUS H30 SDD is dependent on the operating temperature and the shaping time. To achieve good results, KETEK recommends an operating temperature below -20°C. Typical operating conditions are -35°C and an analog shaping time of 4 µs.
The temperature dependency is more significant for longer shaping times.
Finding the optimum operating conditions depends very much on ambient temperature and count rate for the specific application.
The VITUS H30 SDD detector offers a high count rate capability up 500 kcps with weak FWHM dependency as shown in the graph for different analogue shaping times.
The data of the example plot has been acquired with a VITUS H30 Standard Class detector at an operating temperature of -35°C.
The plots show values up to the maximum reasonable dead time of 60%.
| SDD Voltages and Currents | ||
|---|---|---|
| Ring1 (R1) | -20 V ± 5 V | 10 µA typ. |
| RingX (RX) | -130 V ± 20 V | 10 µA typ. |
| Back | -60 V ± 5 V | <1 nA |
| Included FET | ||
|---|---|---|
| Drain | 3 V ± 0.5 V | 3 mA |
| Source | 0 V | |
| Bulk | -5 V ± 3 V | |
| Reset | 1 V | 1 µs |
| Feedback | ramped output | |
| Peltier Element | 4 V | 1 A max |
|---|---|---|
| Temperature Monitor | slope 2.40 mV/K 520 mV @ +20°C |
1 µA |
The VITUS H30 SDD is encapsulated with a bevelled cap for optimum beam geometries in your application. The used TO8 housing can be easily mounted by the M4 stud.
Available versions and ordering codes for Vitus-H30 SDD:
The vacuum type codes end with a 'V' - e.g. 133V